Instruction :
GNL-2100L purity oxygen analyzer is with ion current principle, which is capable of oxygen purity measurement range 10-99.99%.
GNL – 2100L use 3D ion current sensor for measuring unit, through the sensor signal amplifying, filtering, linearization correction after electric processing, through the microprocessor calculated nitrogen oxygen mixed gas concentration of o2, and provides the oxygen concentration. Analyzer with high precision, good stability, long life and other characteristics, suitable for a variety of environments, nitrogen and oxygen mixed gas of oxygen concentration measurement.
Features
- with microprocessor as the core, 3D ion current sensor for measuring unit, with intelligent, good stability, high precision, calibration cycle is long, etc;
- Display intuitive, rich: 128 x 64 LCD screen window shows oxygen concentration, menus, data and equipment working status, display intuitive, rich in content;
- Technology, using the latest touch buttons, easy to operate, buttons with long life;
- Long standby time: built-in high capacity lithium battery, normal use, standby time is not less than 8 hours;
- Change analyzer menu lock switch: to prevent wrong operation parameter affect the normal operation;
- Alarm function: such as oxygen concentration deviates from the value of the set alarm, the analyzer will send out alarm, alarm value can be set up in full scale range.
Measure Range: 0-10/100/1000ppm,0-1%O2
Accuracy: <=±5% in the range of 0-10ppm;s±3% in the range of 0-100ppm;s±2% in the range of 0-100pm;
Display Resolution: 0.01ppm
Response Time: T90<=65s(0-1000ppm)
Measure Medium: H2, He, hydrocarbons, inert gas, mixed gas & acidic gases
Stability: Zero drift <=±1% of FS/7d; Span drift <=±1% of FS/7d
Sample Gas Flow: 800-1000ml/min
Repeatability: <=±2.5% F.S(0-10ppm);<=±1.5% F.S(0-100ppm);<=±1% F.S(0-1000ppm);
Sample Gas Pressure: 35KPa-210KPa
RecoveringTime: Reducing concentration in the air to 1ppm in 60mins
Sensor Life: 24months at 25℃ and 1 atm.
Sensor Principle: Electrochemical Sensor
Warranty: 12 months analyzer; 12 months sensor
Weight: 4.5kg
Appearance dimension: 255mm×250mm×160mm(L×W×H)
It can be used to measure the trace oxygen concentration in the fields of nitrogen separation from air, chemical process, magnetic materials such as high temperature sintering furnace gas protection, the electronics industry protective gas, glass & building materials industries, various mixed gases, etc.