HT-LA433 Oxygen Analyzer
Short Description:
Product Detail
Technical Specifications
Applications
Product Tags
HT-LA433 Oxygen Analyzer
Technical parameters
Measurement Principle: Zirconia |
Display Mode: 320×240 dot matrix color LCD |
Measurement Range: 0~10/100/1000PPm/1.00%/25.00% O2 |
Measurement Accuracy: 0~1000PPm/1.00%/25.00%≤±1%FS0~100PPm≤±2%FS
0~10PPm≤±5%FS |
Resolution: 0.1PPm (trace)0.01% (normal) |
Repeatability: ≤±1.0%FS |
Response Time: T90≤30S |
Output Port: 4-20mA (Non-isolated output, load resistance<1KΩ)0-10V.DC (Non-isolated output, load resistance>10KΩ)
2 ways of programmable alarm output (Dry contact type, maximum load 220V/2A) 1 way of error alarm output |
Communication Mode: RS232 (default) or RS485 |
Power Supply: AC88~264V 50/60Hz, power dissipation<25VA |
Environment Temperature: -10℃~+50℃ |
Storage Environment Humidity: <90%RH, non-condensing |
Operation Environment Humidity: <100%RH, non-condensing |
Sample Gas Temperature: 0℃~60℃ |
Sampling Mode: Gas in |
Sample Gas Flow: 400~600mL/min |
Gas Inlet Pressure: 5Kpa≤ relative pressure≤200Kpa (relief valve is recommended if the pressure >200Kpa) |
Gas Outlet Pressure: Free venting (under safe conditions) |
Background Gas: N2 or other inert gas |
Size: 145mm×175mm×301mm (H×W×D) |
Hole-Cutting Size: 135mm×135mm (H×W) |
Sensor Service Life: >5 years (under normal conditions of use) |
Gas Circuit Port: NPT 1/8 internal thread |
Instrument Weight: Net weight 3.1Kg |
Installation Method: Embedded mounting |
The instrument features
a. Friendly communication menu allows an intuitive and convenient operation; |
b. 320×240 graphic dot-matrix color LCD display is clear to see; |
c. The original imported zirconia sensor consumes no service life when the power is off, and has a high measurement accuracy, rapid response speed and small zero drift value, etc.; |
d. It could operate in air for a long time without shortening sensor’s service life; |
e. The instrument allows multi-point fractional liner calibration, meeting the need for an accurate measurement of oxygen content among the full range; |
f. The response speed is rapid and it could measure 0.1 PPm at minimum; |
g. Automatic temperature and pressure compensation could reduce the impact of environment temperature and pressure changes on the measured accuracy; |
h. 40 thousand pieces of historical data can be stored locally automatically and can be read at any time; |
i. RS232 (default) or RS485 digital communication port can be selected according to clients’ requirement, to realize unidirectional or bidirectional communication with computer or other digital communication facilities. |
Measure Range: 0-10/100/1000ppm,0-1%O2
Accuracy: <=±5% in the range of 0-10ppm;s±3% in the range of 0-100ppm;s±2% in the range of 0-100pm;
Display Resolution: 0.01ppm
Response Time: T90<=65s(0-1000ppm)
Measure Medium: H2, He, hydrocarbons, inert gas, mixed gas & acidic gases
Stability: Zero drift <=±1% of FS/7d; Span drift <=±1% of FS/7d
Sample Gas Flow: 800-1000ml/min
Repeatability: <=±2.5% F.S(0-10ppm);<=±1.5% F.S(0-100ppm);<=±1% F.S(0-1000ppm);
Sample Gas Pressure: 35KPa-210KPa
RecoveringTime: Reducing concentration in the air to 1ppm in 60mins
Sensor Life: 24months at 25℃ and 1 atm.
Sensor Principle: Electrochemical Sensor
Warranty: 12 months analyzer; 12 months sensor
Weight: 4.5kg
Appearance dimension: 255mm×250mm×160mm(L×W×H)
It can be used to measure the trace oxygen concentration in the fields of nitrogen separation from air, chemical process, magnetic materials such as high temperature sintering furnace gas protection, the electronics industry protective gas, glass & building materials industries, various mixed gases, etc.